Document Type

Article

Original Publication Date

2013

Journal/Book/Conference Title

Journal of Transplantation

DOI of Original Publication

10.1155/2013/802618

Comments

Originally published at http://dx.doi.org/10.1155/2013/802618

Date of Submission

August 2014

Abstract

Purpose. Kidneys from DCD donors represent a significant pool, but preservation problems exist. The study objective was to test the importance of machine type for hypothermic preservation of DCD kidneys.Methods. Adult Beagle dog kidneys underwent 45 minutes of warm in situ ischemia followed by hypothermic perfusion for 24 hours (Belzer-MPS Solution) on either an ORS LifePort or a Waters RM3 using standard perfusion protocols. Kidneys were then autotransplanted, and renal function was assessed over 7 days following contralateral nephrectomy. Results. Renal vascular resistance was not different between the two pumps. After 24 hours, the oxygen partial pressure and oxygen delivery in the LifePort perfusate were significantly lower than those in the RM3 but not low enough to change lactate production. TheLifePort ran significantly colder than RM3 (2° versus 5°C). The arterial pressure waveform of the RM3 was qualitatively different from the waveform of the LifePort. Preservation injury after transplantation was not different between the devices. When the LifePort was changed to nonpulsatile flow, kidneys displayed significantly greater preservation injury compared to RM3. Conclusions. Both LifePort and RM3 can be used for hypothermic machine perfusion preservation of DCD kidneys with equal outcomes as long as the duty cycle remains pulsatile.

Rights

Copyright © 2013 Susanne L. Lindell et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Is Part Of

VCU Surgery Publications

Share

COinS