Abstract
Background: Clot formation is the most common mechanical complication of ECMO and can lead to oxygenator failure and the need for subsequent circuit changes. The goals of this study were to identify early indicators of circuit failure to alert providers of ECMO circuit health.
Hypothesis: We hypothesized that patient-specific circuit parameters can predict circuit health to identify risk of early circuit failure in neonate ECMO patients. Using a retrospective chart analysis ECMO flow parameters and clotting factors were identified during the 48 hours prior to ECMO circuit change through the 24 hours post circuit change. Statistical analysis included non-parametric Mann-Whitney U-test.
Results: There was a significant increase in maximum and mean delta-p prior to need for circuit changes compared to those without (p=0.011 and p=0.0128 respectively) and a significant increase in the maximum RPM and mean RPM (p=0.0043 and p=0.0057 respectively). There was a significant increase in mean plasma free hemoglobin (hgb) (p=0.0209); however, the maximum plasma free hgb was not significant (p=0.0569). No differences were notable for sweep and venous pressure in those with circuit changes. Furthermore, clotting parameters were not found to be significant, including ACT, heparin, platelet count, fibrinogen, PT, PTT, INR, AT III (%), anti-Xa.
Conclusion: Changes in Delta-p, RPM, and flow may be valuable predictors of early circuit impairment in neonates on ECMO. Sweep, venous pressure and clotting parameters may not reliable predictors of circuit health.
Publication Date
2021
Keywords
Extracorporeal membrane oxygenation (ECMO), neonatal intensive care unit (NICU), change in pressure, (delta-P), Pump Flow Revolutions per minute (RPM), plasma free hemoglobin (pfHb)